PHOTOACOUSTIC GAS SPECTROSCOPY
USING TUNABLE LIGHT GENERATORS
M. Kaučikasa, Z. Kuprionisb, D. Vasiljeva,
V. Švedasa, V. Vaičikauskasa, and R.
Januškevičiusa
aInstitute of Physics, Laboratory of Nonlinear
Optics and Spectroscopy, Savanorių 231, LT-02300 Vilnius,
Lithuania
bUAB “EKSPLA”, Savanorių 231, LT-02300 Vilnius,
Lithuania
Received 15 December 2004
The performance of phototoacoustic gas
registration using a broad-band optical parametric oscillator
(OPO) was tested. The OPO was capable of generating tunable pulses
of about 7 ns duration and about 5 cm−1 spectral
bandwidth in the 0.72–1.9 μm range (idler), thus covering
overtones of stretching vibrations and combination vibrations of
the hydrogen atom of the analyte molecules. A nonresonant one
microphone acoustic registration scheme was employed. The measured
phototoacoustic spectra of water vapour and methane conform with
absorption spectra calculated with the aid of HITRAN database. For
the 1.361 μm absorption line, the 4.3 ppmv detection
threshold was evaluated for ambient water vapour (ppmv is 1 part
in 106 by volume).
Keywords: photoacoustic gas spectroscopy, pulsed
near-infrared optical parametric oscillator, HITRAN data base
PACS: 43.35.Ud, 82.80.Kq
FOTOAKUSTINĖ DUJŲ SPEKTROSKOPIJA
PANAUDOJANT DERINAMO BANGOS ILGIO ŠVIESOS GENERATORIUS
M. Kaučikasa, Z. Kuprionisb, D. Vasiljeva,
V. Švedasa, V. Vaičikauskasa, R.
Januškevičiusa
aFizikos institutas, Vilnius, Lietuva
bUAB “EKSPLA”, Vilnius, Lietuva
Tikrinamas fotoakustinės dujų registracijos
našumas panaudojant plačios spektrinės juostos optinio
parametrinio osciliatoriaus (OPO) šaltinį. 0,72–1,9 μm
ruože OPO generavo derinamo bangos ilgio maždaug 7 ns trukmės ir
apie 5 cm−1 spektrinės juostos impulsus. Šis spektro
ruožas perkloja analizuojamosios medžiagos molekulių vandenilio
atomų fundamentinių virpesių kombinacinius ir obertonų dažnius.
Naudojama nerezonansinė vieno mikrofono akustinių virpesių
registravimo schema. Pamatuotieji vandens garų ir metano
optoakustiniai spektrai kokybiškai sutampa su šių medžiagų
sugerties spektrais, apskaičiuotais panaudojant HITRAN duomenų
bazę. Registruojant ties 1,361 μm sugerties linija
aplinkos oro vandens garams įvertintas 4,3 ppmv dydžio aptikimo
slenkstis (ppmv yra viena milijoninė tūrinė dalis).
References / Nuorodos
[1] D. Richter, D.G. Lancaster, and F.K. Tittel, Development of an
automated diode-laser-based multicomponent gas sensor, Appl. Opt. 39(24),
4444–4450 (2000),
http://dx.doi.org/10.1364/AO.39.004444
[2] D.G. Lancaster, R. Weidner, D. Richter, F.K. Tittel, and J.
Limpert, Compact CH4 sensor based on difference frequency
mixing of diode lasers in quasi-phasematched LiNbO3, Opt.
Commun. 175(4–6), 461–468 (2000),
http://dx.doi.org/10.1016/S0030-4018(00)00485-5
[3] K.W. Busch and M.A. Busch (eds.), Cavity Ringdown
Spectroscopy (Oxford University Press, New York, 1999),
http://dx.doi.org/10.1021/bk-1999-0720
[4] B.A. Paldus, T.G. Spence, R.N. Zare, J. Oomens, F.J.M. Harren,
D.H. Parker, C. Gmachl, F. Capasso, D.L. Sivco, J.N. Baillargeon,
A.L. Hutchinson, and A.Y. Cho, Photoacoustic spectroscopy using
quantum cascade lasers, Opt. Lett. 24(3), 178–180 (1999),
http://dx.doi.org/10.1364/OL.24.000178
[5] K. Kuhnemann, K. Schneider, A. Hecker, A.A.E. Martis, W. Urban,
S. Schiller, and J. Mlynek, Photoacoustic trace-gas detection using
a cw single-frequency parametric oscillator, Appl. Phys. B 66(6),
741–745 (1998),
http://dx.doi.org/10.1007/s003400050461
[6] D. Hofstetter, M. Beck, J. Faist, M. Nagele, and M.W. Sigrist,
Photoacoustic spectroscopy with quantum cascade distributed feedback
lasers, Opt. Lett. 26(12), 887–889 (2001),
http://dx.doi.org/10.1364/OL.26.000887
[7] L.S. Rothman et al., The HITRAN molecular spectroscopic
database: Edition of 2000 including updates through 2001, J. Quant.
Spectros. Radiat. Transfer 82(1–4), 5–44 (2003),
http://dx.doi.org/10.1016/S0022-4073(03)00146-8
[8] A. Schmohl, A. Miklós, and P. Hess, Detection of ammonia by
photoacoustic spectroscopy with semiconductor lasers, Appl. Opt. 41(7),
1815–1823 (2002),
http://dx.doi.org/10.1364/AO.41.001815